Results from the national intercomparison for rise time and bandwidth measurements within the German Calibration Service (Deutscher Kalibrierdienst – DKD)

Kai Baaske, Thomas Kleine-Ostmann, and Thorsten Schrader

Viewed

Total article views: 902 (including HTML, PDF, and XML)
HTML PDF XML Total BibTeX EndNote
511 328 63 902 56 69
  • HTML: 511
  • PDF: 328
  • XML: 63
  • Total: 902
  • BibTeX: 56
  • EndNote: 69
Views and downloads (calculated since 19 Sep 2019)
Cumulative views and downloads (calculated since 19 Sep 2019)

Viewed (geographical distribution)

Total article views: 830 (including HTML, PDF, and XML) Thereof 825 with geography defined and 5 with unknown origin.
Country # Views %
  • 1
1
 
 
 
 

Cited

Latest update: 14 Dec 2024
Download
Short summary
An intercomparison in the framework of the German Calibration Service (Deutscher Kalibrierdienst – DKD) was carried out. Two different travelling standards were used to measure the rise time of a pulse generator and the rise time, bandwidth and attenuation of an oscilloscope. Fourteen participants performed the measurements from September 2015 until May 2016. The overall results show that the measurements of the participating laboratories are in good agagreement with the reference values.